arXiv · 2212.04773
Pixelated silicon detector for radiation beam profile measurements
Abstract
A pixelated silicon detector, developed originally for particle physics experiments, was used for a beam profile measurement of a Co-60 irradiator in a water phantom. The beam profile was compared to a profile measured with a pinpoint ionization chamber. The differences in the pixel detector and pinpoint chamber relative profiles were within approximately 0.02, and after calculating correction factors with Monte Carlo simulations for the pixel detector, the differences were decreased to almost less than 0.005. The detector's capability to measure pulse-height was used to record an electron pulse-height spectrum in water in the Co-60 beam, and the results agreed well with simulations.
Explore related subjects
Keep this discovery
Joonas Tikkanen, Stefanie Kirschenmann, Nikita Kramarenko, Erik Brücken, Pirkitta Koponen, Panja Luukka, Teemu Siiskonen, Raimo Turpeinen, Jennifer Ott. 2022-12-09. Pixelated silicon detector for radiation beam profile measurements. https://arxiv.org/abs/2212.04773
Cite the original work for its findings. Save a collection to share your selection of sources.