Inkwell: Design and Validation of a Low-Cost Open Electricity-Free 3D Printed Device for Automated Thin Smearing of Whole Blood

Abstract

Microscopy plays a crucial role in hematology and diagnosis of infectious diseases worldwide. For malaria alone, more than 200 million slides are read by manual microscopists every year. High quality thin blood smears are essential for subsequent microscopy examinations including malaria microscopy, but are hard to make in field settings. Existing devices for assisting in making thin smears are available but are limited by cost or complexity for wider use. Here we present Inkwell, a portable mechanical device capable of making high quality thin blood smears in field settings. Inkwell is simple, low-cost, does not use electricity, and requires minimal training prior to use. By utilizing passive dissipative dynamics of a spiral spring coupled to an air dashpot with a tunable valve - we demonstrate a highly tunable mechanism for constant velocity smears at prescribed angle. Inkwell is capable of producing high quality blood smears of tunable cell density with more than 12 million individually distinguishable red blood cells on a single slide. The current design, which exploits precision manufacturing of a 17 cents plastic syringe and a spring, can be printed on a standard 3D printer with overall unit cost of less than a few dollars in large quantities. We further present usability tests to confirm performance over 10,000 unit cycle operations with no degradation in quality of the smear and demonstrate ease of use with minimal training. Inkwell enhances the broader toolbox of open innovations in diagnostics for providing high quality medical care in low and medium resource settings. Combined with rise of 3D printing, Inkwell presents an alternative to traditional centralized manufacturing and opens up distributed manufacturing of medical diagnostics in global context.

Explore related subjects

Keep this discovery

BibTeXRIS

Jerome Nowak, Anesta Kothari, Hongquan Li, Jaspreet Pannu, Dani Algazi, Manu Prakash. 2023-04-20. Inkwell: Design and Validation of a Low-Cost Open Electricity-Free 3D Printed Device for Automated Thin Smearing of Whole Blood. https://arxiv.org/abs/2304.10200

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph