arXiv · 2401.08701
In-silico Analysis of Hemodynamic Indicators in Idealized Stented Coronary Arteries for Varying Stent Indentation
Abstract
In this work, we investigate the effects of stent indentation on hemodynamic indicators in stented coronary arteries. Our aim is to assess in-silico risk factors for in-stent restenosis (ISR) and thrombosis after stent implantation. The proposed model is applied to an idealized artery with Xience V stent for four indentation percentages and three mesh refinements. We analyze the patterns of hemodynamic indicators arising from different stent indentations and propose an empirical frequency analysis of time-averaged WSS (TAWSS), oscillatory shear index (OSI), and relative residence time (RRT). We observe that higher indentations display higher frequency of critically low TAWSS and non-physiological OSI and RRT. Furthermore, an appropriate mesh refinement is needed for accurate representation of hemodynamics in the stent vicinity. The results provide physics-based evidence for the correlation between high indentation and ISR.
Explore related subjects
Keep this discovery
Anna Ranno, Kiran Manjunatha, Anne Glitz, Nicole Schaaps, Stefanie Reese, Felix Vogt, Marek Behr. 2024-01-14. In-silico Analysis of Hemodynamic Indicators in Idealized Stented Coronary Arteries for Varying Stent Indentation. https://doi.org/10.1080/10255842.2024.2382819
Cite the original work for its findings. Save a collection to share your selection of sources.