High-count-rate Particle Tracking in Proton and Carbon Radiotherapy with Timepix2 Operated in Ultra-Short Acquisition Time

Abstract

This work investigates the operational acquisition time limits of Timepix3 and Timepix2 detectors operated in frame mode for high-count rate of high deposited energy transfer particles. Measurements were performed using alpha particles from a 241Am laboratory source and proton and carbon ion beams from a synchrotron accelerator. The particle count rate upper limit is determined by overlapping per-pixel particle signals, identifiable by the hits per pixel counter > 2, indicating the need to decrease acquisition time. On the other hand, the lower limit is the time required to collect the particle deposited charge while maintaining spectral properties. Different acquisition times were evaluated for an AdvaPIX Timepix3 detector (500 um Silicon sensor) with standard per-pixel DAC settings and a Minipix Timepix2 detector (300 um Silicon sensor) with standard and customized settings the pulse shaping parameter and threshold. For AdvaPIX Timepix3, spectra remained accurate down to 100 us frame acquisition time; at 10 us, loss of collected charge occurred, suggesting either avoiding this acquisition time or applying a correction. Timepix2 allowed acquisition times down to 100 ns for single particle track measurements even for high energy loss, enabled by a new Timepix2 feature delaying shutter closure until full particle charge collection. This work represents the first measurement utilizing Timepix-chips pixel detectors in an accelerator beam of clinical energy and intensity without the need to decrease the beam current. This is made possible by exploiting the short shutter feature in Timepix2 and a customized per-pixel energy calibration of the Timepix2 detector with a larger discharging signal value which allowed for a shorter time-over-threshold (ToT) signal. These customized settings extend the operation of the pixel detectors to higher event rates up to 10^9 particles/cm^2/s.

Explore related subjects

Keep this discovery

BibTeXRIS

C. Oancea, A. Resch, S. Barna, G. Magrin, L. Grevillot, D. Hladik, L. Marek, J. Jakubek, C. Granja. 2024-09-26. High-count-rate Particle Tracking in Proton and Carbon Radiotherapy with Timepix2 Operated in Ultra-Short Acquisition Time. https://arxiv.org/abs/2409.17660

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det