Ungated, plug-and-play preclinical cardiac CEST-MRI using radial FLASH with segmented saturation

Abstract

Purpose: ECG and respiratory-gated preclinical cardiac CEST-MRI acquisitions are difficult due to variable saturation recovery with T1, RF interference in the ECG signal, and offset-to-offset variation in Z-magnetization and cardiac phase introduced by changes in cardiac frequency and trigger delays. Methods: The proposed method consists of segmented saturation modules with radial FLASH readouts and golden angle progression. The segmented saturation blocks drive the system to steady-state, and because center k-space is sampled repeatedly, steady-state saturation dominates contrast during gridding and reconstruction. Ten complete Z-spectra were acquired in healthy mice using both ECG and respiratory-gated, and ungated methods. Z-spectra were also acquired at multiple saturation B1 values to optimize for amide and creatine contrasts. Results: There was no significant difference between CEST contrasts (amide, creatine, MT) calculated from images acquired using ECG and respiratory-gated and ungated methods (p = 0.27, 0.11, 0.47). A saturation power of 1.8$\mu$T provides optimal contrast amplitudes for both amide and total creatine contrast without significantly complicating CEST contrast quantification due to water direct saturation, magnetization transfer, and RF spillover between amide and creatine pools. Further, variability in CEST contrast measurements was significantly reduced using the ungated radial FLASH acquisition (p = 0.002, 0.006 for amide and creatine respectively). Conclusion: This method enables CEST mapping in the murine myocardium without the need for cardiac or respiratory gating. Quantitative CEST contrasts are consistent with those obtained using gated sequences, and per-contrast variance is significantly reduced. This approach makes preclinical cardiac CEST-MRI easily accessible, even for investigators without prior experience in cardiac imaging.

Explore related subjects

Keep this discovery

BibTeXRIS

Jonah Weigand-Whittier, Michael Wendland, Bonnie Lam, Mark Velasquez, Moriel H. Vandsburger. 2024-11-01. Ungated, plug-and-play preclinical cardiac CEST-MRI using radial FLASH with segmented saturation. https://doi.org/10.1002/mrm.30382

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph