Proposal and Evaluation of a Practical CBCT Dose Optimization Method
Abstract
Objective: To propose a CBCT dose optimization method using readily available measurement equipment in radiation oncology departments. Approach: A 0.6cc air kerma (Kair) calibrated Farmer chamber measured Kair at isocenter for five default CBCT protocols (HEAD, PELVIS, LARGE PELVIS, THORAX, SPOTLIGHT) on a Varian TrueBeam linac. Imaging parameters were modified to reduce cumulative exposure by reducing projections (by 25%, 50%, 75%) in FDK reconstruction, or lowering single frame exposure by 25%, 50%, 70 to 75%. Cone Beam Dose Index (CBDI) and weighted CBDI (CBDI_w) were measured in CTDI phantoms to correlate with Kair. Image quality was assessed quantitatively (HU linearity, uniformity, low-contrast visibility, high-contrast resolution, spatial integrity) using Catphan604, and qualitatively via anthropomorphic phantoms. Results: A linear relationship between Kair, CTDI, and CBDI_w was established. High-contrast resolution and spatial integrity were unaffected by dose reduction (R < 0.5), while low-contrast visibility strongly correlated with exposure (R > 0.75). HU uniformity and linearity remained stable (<40 HU deviation), except for LARGE PELVIS at <50% single-frame exposure. 80 to 50% dose reduction retained clinically acceptable image quality with minimal artifacts or noise. Significance: CBCT protocols on TrueBeam linacs were optimized for dose reduction. The method enables easy implementation by radiotherapy physicists without diagnostic physics equipment or expertise.
Explore related subjects
Keep this discovery
S. Gros, J. Bian, J. Jackson, M. Delafuente, H. Kang, W. Small Jr, M. Mahesh. 2025-02-03. Proposal and Evaluation of a Practical CBCT Dose Optimization Method. https://arxiv.org/abs/2502.01509
Cite the original work for its findings. Save a collection to share your selection of sources.