Ultrasound imaging of cortical bone: cortex geometry and measurement of porosity based on wave speed for bone remodeling estimation

Abstract

Intracortical US imaging extends B-mode imaging into bone using a dedicated image reconstruction algorithm that corrects for refraction at the bone-soft tissue interfaces. It has shown promising results in a few healthy, predominantly young adults, providing anatomical images of the cortex (periosteal and endosteal surfaces) along with estimations of US wave speed. However, its reliability in older or osteoporotic bones remains uncertain. In this study, we critically assessed the performance of intracortical US imaging ex vivo in bones with various microstructural patterns, including bones exhibiting signs of unbalanced intracortical remodeling. We analyzed factors influencing US image quality, particularly endosteal surface reconstruction, as well as the accuracy of wave speed estimation and its relationship with porosity. We imaged 20 regions of interest from the femoral diaphysis of five elderly donors using a 2.5 MHz US transducer. The reconstructed US images were compared to site-matched high-resolution micro-CT (HR-muCT) images. In samples with moderate porosity, the endosteal surface was accurately identified, and thickness estimates from US and HR-muCT differed by less than 10%. In highly remodeled bones with increased porosity, the reconstructed endosteal surface appeared less bright and was located above the cortex region containing resorption cavities. We observed a decrease in US wave speed with increasing cortical porosity suggesting that the method could discriminate between bones with low porosity (less than 5%) and those with moderate to high porosity (greater than ~10%). This study paves the way for the application of US imaging in diagnosing cortical bone health, particularly for detecting increased cortical porosity and reduced cortical thickness.

Explore related subjects

Keep this discovery

BibTeXRIS

Amadou S. Dia, Guillaume Renaud, Christine Chappard, Quentin Grimal. 2025-02-12. Ultrasound imaging of cortical bone: cortex geometry and measurement of porosity based on wave speed for bone remodeling estimation. https://arxiv.org/abs/2502.08824

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph