Terahertz frequency-domain 4x4 Mueller matrix ellipsometer instrument designed for high-frequency magnetic resonance measurements

Abstract

We report a Mueller matrix ellipsometer design using dual continuously rotating anisotropic meta wave plates which determines the full set of Mueller matrix elements in the terahertz spectral range. The instrument operates in the frequency domain and employs a frequency tunable, solid state synthesizer based, continuous wave terahertz source with sub-MHz bandwidth. The implemented source permits operation within 82-125 GHz and 170-250 GHz, without and with an additional frequency doubler, respectively. The terahertz transparent meta wave plates consist of 3D-printed polymer based columnar thin film structures with subwavelength dimensions. The rotating wave plates produce sufficient modulation of the Stokes vector components of transmitted terahertz light to use the wave plates in polarization state generator and polarization state detector devices. Fast terahertz light detection rate of a quasi optical solid state detector permits acquisition with few microsecond temporal resolution and electronic sweeping control of the source frequency. We develop a fast frequency sweeping scheme while continuously rotating the terahertz wave plates. Subsequent sorting of measured data permits measurements of Mueller matrix elements at hundreds of different wavelengths. The Mueller matrix elements are obtained by forward numerical reduction of the measured data using an algorithm developed by Ruder et al.[Opt. Lett. 45, 3541 (2020)]. The instrument is combined with a magnetocryostat capable of reaching magnetic field strengths at the sample position of -8 T...8 T and sample temperatures from 4 K...400 K. Hence, the instrument is suitable to measure the full Mueller matrix of samples with magnetic resonances as demonstrated recently by Rindert et al. [Phys. Rev. B 110, 054413 (2024)]. We discuss design, calibration, and example applications.

Explore related subjects

Keep this discovery

BibTeXRIS

Viktor Rindert, Alexander Ruder, Steffen Richter, Philipp Kühne, Mark Bauer, Vanya Darakchieva, Mathias Schubert. 2025-03-28. Terahertz frequency-domain 4x4 Mueller matrix ellipsometer instrument designed for high-frequency magnetic resonance measurements. https://arxiv.org/abs/2503.22500

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det