High-Resolution Directional Depth Electrodes: Open-Source FEM Lead-Field Modeling, Characterization, and Validation

Abstract

Depth electrodes used in stereoelectroencephalography (sEEG) and deep-brain stimulation (DBS) are essential tools for neural recording and stimulation. Traditional designs have limited spatial resolution, typically 8 to 16 cylindrical contacts (0.8 to 1.0 mm diameter) along a 5 to 10 cm shaft, restricting recordings from small or localized populations. Recent high-density, directional electrodes (HDsEEG) enable finer localization of local field potentials (LFPs) and spike timing. Yet, characterizing their directional sensitivity and validating modeling tools for lead field (LF) analysis remain critical. We compare finite element method (FEM) LF modeling of a novel HD-sEEG electrode using two tools: a commercial solver (ANSYS) and an open-source pipeline (Brainstorm-DUNEuro). Goals: (i) validate against analytical solutions, (ii) assess solver differences, and (iii) characterize HDsEEG directional sensitivity. LFs were modeled in simple and bio-relevant scenarios. Using Helmholtz reciprocity, we computed LFs by (a) electrode-based stimulation with ANSYS and (b) source-based recording with Brainstorm-DUNEuro. First, a multi-sphere head model with known solutions tested the solver's accuracy. Next, an HDsEEG electrode in a homogeneous conductor was simulated. Directional effects were assessed by comparing sensitivity with vs. without the insulating substrate. Source localization performance was also compared between HDsEEG and standard electrodes. Both solvers closely matched analytic solutions. In realistic settings, LF distributions were highly similar. Modeling showed clear directional sensitivity: contacts facing a source had higher sensitivity than those shadowed, reflecting a substrate shielding effect that vanished when the substrate was conductive. Crucially, HDsEEG improved source localization, as voltage differences across contacts provided robust directional LF information.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Takfarinas Medani, Jace Willis, Chris Wright, Yash Vakilna, Ryan Shores, Raymundo Cassani, Anand Joshi, Richard Leahy, John Seymour, John Mosher. 2025-08-16. High-Resolution Directional Depth Electrodes: Open-Source FEM Lead-Field Modeling, Characterization, and Validation. https://arxiv.org/abs/2508.13212

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph