Quantitative In Vivo Cherenkov Luminescence Imaging and Dosimetry of Yttrium-86-NM600

Abstract

Purpose: The expansion of radiopharmaceutical therapy (RPT) development demands scalable preclinical dosimetry methods. While PET and SPECT remain the gold standards, their low throughput and high cost limit large-cohort studies. Cherenkov luminescence imaging (CLI) offers a high-throughput alternative but suffers from depth-dependent attenuation and photon scatter that compromise quantitative accuracy. This work develops and validates a quantitative CLI methodology incorporating attenuation and scatter corrections for accurate preclinical dosimetry. Methods: Depth-dependent attenuation was characterized using a tissue-mimicking phantom to derive calibration coefficients. Photon scatter was modeled using GEANT4-generated Cherenkov spread functions (CSFs), applied in a depth-weighted iterative Richardson--Lucy deconvolution/reconvolution framework. The method was evaluated in NU/NU mice (n=4) bearing MC38 tumors after injection of $^{86}$Y-NM600, an isotope suitable for both PET and CLI. Liver and tumor activities were quantified at four timepoints using PET and the proposed CLI method. Monte Carlo dosimetry was performed for both modalities. Results: CLI--PET activity quantification yielded mean errors of 15.4% (liver) and 10.3% (tumor) over the first three timepoints. Tumor absorbed doses from CLI-derived synthetic PET images (3.4 $\pm$ 0.3 Gy/MBq) were statistically indistinguishable from PET-based estimates (3.2 $\pm$ 0.2 Gy/MBq, $p=0.31$). Discrepancies increased at late timepoints due to low activity and background auto-luminescence. Conclusions: With appropriate depth-dependent attenuation calibration and Monte Carlo--derived scatter correction, CLI can provide quantitative biodistribution and dosimetry estimates comparable to PET. This approach enables high-throughput, low-cost in vivo dosimetry, expanding the feasibility of large-scale preclinical RPT studies.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Campbell Haasch, Malick Bio Idrissou, Sydney Jupitz, Aubrey Parks, Reinier Hernandez, Brian Pogue, Bryan Bednarz. 2025-09-16. Quantitative In Vivo Cherenkov Luminescence Imaging and Dosimetry of Yttrium-86-NM600. https://arxiv.org/abs/2509.13533

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph