Comparative study of YSO, GaGG, and BGO scintillators coupled to a SiPM array for gamma-ray spectroscopy
Abstract
A compact gamma-ray detector module was developed and characterized for spectroscopic applications across a wide energy range. The detector comprises a 4x4 array of silicon photomultipliers (SiPMs, Hamamatsu MPPC S13360-3050) coupled with three different inorganic scintillators (YSO, GaGG, BGO). The performance of the detector was evaluated over an energy range from 32 keV to 2500 keV using various standard gamma-ray point sources ({241}Am, {137}Cs, {22}Na, {60}Co, {228}Th). All three detector configurations demonstrated linearity between pulse height and gamma-ray energy over most of the tested range. GAGG exhibited the most consistent linear behavior up to 2500 keV. YSO with a faster decay constant showed linearity up to ~2000 keV, followed by slight saturation effects likely due to SiPM microcell occupancy limits at higher photon flux. BGO, characterized by a low light output and longer decay time, demonstrated stable linearity up to 2500 keV but reduced sensitivity below 300 keV. The measured energy resolution at 662 keV was 8.17% for GAGG, 9.3% for YSO, and 10.2% for BGO, consistent with their light yield and photon detection efficiency (PDE) - matching characteristics. The obtained results showed the trade-offs between scintillator materials in terms of energy resolution, dynamic range, and linearity, and confirm the suitability of the tested configurations for compact gamma-ray spectroscopy applications across the studied energy range.
Explore related subjects
Keep this discovery
S. Nuruyev, G. Ahmadov, D. Berikov, M. Holik, F. Ahmadov, A. Sadigov, R. Akbarov, J. Naghiyev, I. Nuruyev, A. Mammadli. 2025-09-30. Comparative study of YSO, GaGG, and BGO scintillators coupled to a SiPM array for gamma-ray spectroscopy. https://arxiv.org/abs/2509.26252
Cite the original work for its findings. Save a collection to share your selection of sources.