SearcharxivSearch

arXiv · 2510.14437

Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis

Abstract

A systematic study is conducted to understand the coincident resolving time (CRT) for a pair of Lutetium-yttrium oxyorthosilicate (LYSO) and the plastic scintillation detector bars under the Geant4 framework. Crystals are coupled to a silicon photomultiplier single pad wafer with an appropriate optical coupling for signal readout. Pad reads the light photons undergoing optoelectronic conversion at the wafer site and generates electrical pulses with a bi-exponential shape. These signals are used to determine the trigger time stamp of back-to-back gamma-rays emitted from a point source, enabling the evaluation of CRT performance at different plastic scintillator lengths. For the LYSO detector, the simulation yields the CRT response of 174 ps, which aligns to the literature reported value for the dimensions 2 mm x 2 mm x 10 mm. To identify the plastic scintillator dimensions with an integrated gamma-ray detection efficiency comparable to LYSO's photopeak efficiency at 511 keV gamma photons, various bar lengths of commercial plastic BC404 and TIFR Ooty's in-house developed plastic material are attempted in Geant4. Consequently, for both the plastic scintillators, the equivalent length (for the same cross-sectional area) was found to be 4 times that of LYSO crystal length at a threshold of 25 keV. CRT value determined for this dimension is found to be $\approx$ 300 ps for both the plastic medium. It suggests that, if an animal preclinical PET scanner is developed with plastic bars, the minimum achievable image resolution (FWHM) of $\approx$ 4.5 cm can be expected for a pair of detection elements.

Explore related subjects

Keep this discovery

BibTeXRIS

Davinder Siwal, P. K. Mohanty, D. Bose, A. Jain. 2025-10-16. Evaluating plastic scintillator performance as a substitute of LYSO in SiPM based animal PET scanners: A GEANT4 simulation analysis. https://arxiv.org/abs/2510.14437

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det