Laser-Induced Fluorescence for measuring salt mass fraction in evaporating saline solutions

Abstract

A non-intrusive optical diagnostic based on Laser-Induced Fluorescence (LIF) is developed for measuring salt mass fraction in highly concentrated aqueous solutions relevant to absorption heat pump applications. The method relies on a systematic redshift of the fluorescence emission spectra of several dyes, likely induced by an increase of ionic strength with salt mass fraction. A two-color ratiometric approach, based on the ratio of fluorescence intensities in two spectral bands, enables robust mass fraction measurements. This study demonstrates that the techniques is not limited to LiBr but also NaCl, relevant for desalination applications. In parallel, fluorescence lifetime measurements using Time-Correlated Single Photon Counting (TCSPC) reveals that fluorescence lifetime is also affected by the LiBr mass fraction, allowing an alternative measurement technique. For both techniques, many dyes are tested and are compared in terms of salt mass fraction sensitivity. Both techniques are validated independently through comparison with shadowgraphbased measurements during evaporation of droplets. Good agreement is obtained in liquid-phase conditions over a wide range of salt mass fractions, confirming the reliability of the proposed diagnostics. Key sources of uncertainty including temperature, reabsorption effects and phase changes has been identified and analyzed. Finally, preliminary results demonstrate the potential extension of the approach toward planar imaging of evaporating droplets and more complex multiphase flows. The methodology provides a basis for future quantitative diagnostics of heat and mass transfer in desorption systems such as falling-films.

Explore related subjects

Keep this discovery

BibTeXRIS

Antoine Dumain, Florine Giraud, Ophélie Caballina, Benoit Stutz, Guillaume Castanet. 2026-07-10. Laser-Induced Fluorescence for measuring salt mass fraction in evaporating saline solutions. https://arxiv.org/abs/2607.09220

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det