arXiv · 2607.22870
Commissioning and first results from the Cold Radon Emanation Facility
Abstract
Radon emanation from detector materials is a critical background for next-generation rare event searches, in particular those using noble liquid targets. While highly sensitive screening facilities mitigate this risk prior to detector construction, room-temperature assays often fail to predict cold emanation rates due to temperature-dependent diffusion suppression. The Cold Radon Emanation Facility at Rutherford Appleton Laboratory addresses this by performing high-sensitivity assays at detector operating temperatures. It includes a 2.7 L small-sample chamber, a 200 L chamber for large as-built components operated with a radon concentration line, a cryogenic infrastructure enabling measurements of emanation as a function of temperature, and an electrostatic radon detector, which achieves a minimum detectable activity of ~0.05 mBq at 90% CL. Commissioning results and initial comparative assays are reported, including a preliminary indication of a factor of ~2 suppression of $^{222}$Rn emanation in a titanium sample at cryogenic temperatures. This result, obtained as part of commissioning measurements, illustrates the potential impact of temperature-dependent effects and underscores the importance of in-situ cold assays for future noble liquid detector components.
Explore related subjects
Keep this discovery
Emily Perry, Chamkaur Ghag, Maurits Van der Grinten, Ruben Saakyan, Andrew Stevens, Mark Tucker, Umit Utku. 2026-07-24. Commissioning and first results from the Cold Radon Emanation Facility. https://arxiv.org/abs/2607.22870
Cite the original work for its findings. Save a collection to share your selection of sources.