Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy

Abstract

High-dose-rate (HDR) brachytherapy is highly conformal, but its accuracy depends on catheter and source placement, and clinical workflows lack real-time in vivo dosimetry (IVD) that verifies delivery without disrupting the procedure. We characterize a scintillator-based optical dosimeter integrated into a transrectal probe that tracks source position, speed, dose rate, and time-resolved cumulative dose. A 67 mm x 5 mm line scintillator is imaged by a CMOS camera through a 30-degree endoscopic lens; geometric and flat-field corrections remove barrel distortion and vignetting. Absolute dose was calibrated with a TOPAS Monte Carlo model of the probe's air-cavity geometry. The system was validated in a water phantom with an Ir-192 source and needles at radial offsets of 7, 10, and 15 mm. Response varied by $\pm$2\% along the 67 mm scintillator and was linear over 5-21 cGy/s ($R^2 \geq 0.995$). The Monte Carlo model showed a 3.75\% dose-rate reduction relative to TG-43 water-equivalent conditions. In phantom, the system resolved a 3 mm needle displacement, confirmed by CT. Measured dose-rate profiles agreed with Monte Carlo in peak magnitude and spatial distribution, and integral dose agreed within 1\%. By mapping dose directly to the measurement plane of a standard ultrasound transducer, this system provides high-resolution, time-resolved verification of HDR delivery and a clinical tool for detecting delivery errors with minimal changes to existing treatment protocols.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Kevin Willy, Megan Clark, Petr Bruza, Shauna McVorran, David Gladstone. 2026-11-20. Optical Scintillation Imaging for Dynamic Source Tracking Dosimetry During Prostate HDR Brachytherapy. https://searcharxiv.com/abs/2611.27146

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Searcharxiv by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related papers

Physics-Based Differentiable Beamforming for Deep Learning-Based Contrast Enhanced Ultrasound Super-Resolution Imaging

Ultrasound is widely used for vascular imaging because it is non-invasive, real-time, and cost-effective. However, its spatial resolution remains limited by diffraction to an order of the acoustic wavelength. This limitation is particularly restrictive for imaging of the microvasculature and high-resolution vector-flow-imaging. In this work, we present a multi-channel physics-based deep learning approach to super-resolve monodisperse microbubbles directly from multi-channel radio-frequency (RF) data. We also present a novel image-based metric for model evaluation. The proposed pipeline combines a learnable deconvolution network with a differentiable beamforming layer, enabling a direct deconvolution of the microbubble (MB) signature prior to image formation. By operating on multi-channel RF data, the method aims to improve localization robustness in the presence of dense microbubble populations and deconvolve the lateral direction in a relatively low-frequency regime (1.7 MHz) where MBs exhibit nonlinear behavior. The model is trained and evaluated on simulated data using monodisperse lipid-coated MBs over a broad range of acoustic pressures and microbubble densities. The proposed method improves the signal-to-noise ratio (SNR) compared to the single-channel approach. It achieves sub-wavelength axial and lateral resolution while slightly improving robustness to dense microbubble populations outside the training domain. It also demonstrates generalizability across transmit configurations without requiring explicit training on each transmit configuration.

physics.med-ph↗

Geometry-bridge flow matching for blind rigid-motion correction in head cone-beam CT

Rigid motion of a patient head during head cone-beam CT (CBCT) scan can degrade image quality and complicate recovery of accurate anatomical structures from the measured projections. We present a 3D flow-matching prior approach that learns the image changes induced by motion. Its geometry-bridge connects motion-corrupted and motion-free Feldkamp-Davis-Kress (FDK) reconstructed images of the same source volume through progressively reduced simulated motion. The forward projector's geometry derivative supplies analytic velocity targets along this path. A plug-and-play loop encompasses the learned image updates together with spline-based pose estimation, conjugate-gradient data consistency, and image total-variation regularization. Both the refined image volume and the FDK reconstructed image volume at the estimated poses are evaluated. On simulated 360-view scans from 30 held-out CQ500 patients with nominal per-component motion amplitudes within 10 mm / 10 degrees, the framework achieved state-of-the-art image reconstruction quality and motion accuracy in comparison with other methos such as learned autofocus and diffusion-based joint reconstruction. Mean peak signal-to-noise ratio (PSNR) reached 37.1 dB and mean detector-plane reprojection error (RPE) was 0.27 mm. With matched endpoints at the inference settings, the geometry bridge improved PSNR by 1.6 dB and reduced RPE by 33% relative to image-linear training. With 50 prior updates and coarse-to-fine motion estimation, total processing time averaged 9.1 minutes per patient on an RTX A6000.

physics.med-ph↗

Three-Dimensional Ultrasound Localization Microscopy Using Hadamard-Encoded Multipulse and 4D Denoising on a 256-Channel System with a 1024-element Matrix Array: A Feasibility Study in Human Kidney Transplants

Three-dimensional ultrasound localization microscopy (3D ULM) enables volumetric visualization of microvasculature beyond the acoustic diffraction limit, but its clinical translation remains limited by long acquisition times, low microbubble (MB) signal-to-noise ratio (SNR), and insufficient MB contrast, particularly on ultrasound systems with a limited number of acquisition channels. In this study, we developed an in-human 3D ULM framework for a 256-channel research ultrasound system equipped with a 1024-element matrix array. The framework combines diverging-wave Hadamard-encoded multipulse (HEM) transmission to enhance MB SNR with denoising strategy combining MB enhancement and four-dimensional non-local means filtering to improve MB contrast. The framework was implemented on a research ultrasound platform and evaluated in phantom experiments and four in vivo human kidney transplants. In phantom experiments, HEM improved the MB contrast ratio around 2 dB, while the proposed denoising strategy enhanced in vivo MB contrast by up to around 9 dB. 3D ULM reconstruction achieved a spatial resolution of approximately 84 um, as estimated by Fourier shell correlation. The number of reconstructed microvascular voxels increased exponentially with acquisition duration, reaching approximately 74% of the estimated saturated vessel density using 3000 volumes (~12 s). The subject with the highest histopathological injury exhibited quantitatively reduced microvascular vessel density, flow velocities, and mean perfusion index. These results demonstrate that the proposed framework enables high-quality volumetric super-resolution imaging and quantitative characterization of renal microvasculature using a 256-channel ultrasound system, providing a practical step toward the clinical translation of in vivo 3D ULM.

physics.med-ph↗