arXiv · astro-ph/0508554
Hectospec, the MMT's 300 Optical Fiber-Fed Spectrograph
Abstract
The Hectospec is a 300 optical fiber fed spectrograph commissioned at the MMT in the spring of 2004. A pair of high-speed six-axis robots move the 300 fiber buttons between observing configurations within ~300 s and to an accuracy ~25 microns. The optical fibers run for 26 m between the MMT's focal surface and the bench spectrograph operating at R~1000-2000. Another high dispersion bench spectrograph offering R~5,000, Hectochelle, is also available. The system throughput, including all losses in the telescope optics, fibers, and spectrograph peaks at ~10% at the grating blaze in 1" FWHM seeing. Correcting for aperture losses at the 1.5" diameter fiber entrance aperture, the system throughput peaks at $\sim$17%. Hectospec has proven to be a workhorse instrument at the MMT. Hectospec and Hectochelle together were scheduled for 1/3 of the available nights since its commissioning. Hectospec has returned \~60,000 reduced spectra for 16 scientific programs during its first year of operation.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Daniel Fabricant, Robert Fata, John Roll, Edward Hertz, Nelson Caldwell, Thomas Gauron, John Geary, Brian McLeod, Andrew Szentgyorgyi. 2005-08-25. Hectospec, the MMT's 300 Optical Fiber-Fed Spectrograph. https://doi.org/10.1086/497385
Cite the original work for its findings. Save a collection to share your selection of sources.