arXiv · hep-ex/9504010
A liquid-helium cooled large-area silicon PIN photodiode x-ray detector
Abstract
An x-ray detector using a liquid-helium cooled large-area silicon PIN photodiode has been developed along with a tailor-made charge sensitive preamplifier whose first-stage JFET has been cooled. The operating temperature of the JFET has been varied separately and optimized. The x- and $γ$-ray energy spectra for an \nuc{241}{Am} source have been measured with the photodiode operated at 13 K. An energy resolution of 1.60 keV (FWHM) has been obtained for 60-keV $γ$ rays and 1.30 keV (FWHM) for the pulser. The energy threshold could be set as low as 3 keV. It has been shown that a silicon PIN photodiode serves as a low-cost excellent x-ray detector which covers large area at 13 K.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Yoshizumi Inoue, Shigetaka Moriyama, Hideyuki Hara, Makoto Minowa, Fumio Shimokoshi. 1995-04-27. A liquid-helium cooled large-area silicon PIN photodiode x-ray detector. https://doi.org/10.1016/0168-9002(95)00872-1
Cite the original work for its findings. Save a collection to share your selection of sources.