arXiv · physics/0602159
Characterization of a CCD array for Bragg spectroscopy
Abstract
The average pixel distance as well as the relative orientation of an array of 6 CCD detectors have been measured with accuracies of about 0.5 nm and 50 $μ$rad, respectively. Such a precision satisfies the needs of modern crystal spectroscopy experiments in the field of exotic atoms and highly charged ions. Two different measurements have been performed by illuminating masks in front of the detector array by remote sources of radiation. In one case, an aluminum mask was irradiated with X-rays and in a second attempt, a nanometric quartz wafer was illuminated by a light bulb. Both methods gave consistent results with a smaller error for the optical method. In addition, the thermal expansion of the CCD detectors was characterized between -105 C and -40 C.
Explore related subjects
Keep this discovery
Paul Indelicato, Eric-Olivier Le Bigot, Martino Trassinelli, Detlev Gotta, Maik Hennebach, Nick Nelms, Christian David, Leopold M. Simons. 2006-02-23. Characterization of a CCD array for Bragg spectroscopy. https://doi.org/10.1063/1.2194485
Cite the original work for its findings. Save a collection to share your selection of sources.