arXiv · 0709.3481
Energy resolution improvement in room-temperature CZT detectors
Abstract
We present methods to improve the energy resolution of single-channel, room-temperature Cadmium-Zinc-Telluride (CZT) detectors. A new preamplifier design enables the acquisition of the actual transient current from the crystals and straightforward data analysis methods yield unprecedented energy resolution for our test-detectors. These consist of an eV-CAPture Plus crystal as standard and 1 cm cube Frisch collar crystals created in-house from low-grade coplanar grid detectors. Energy resolutions of 1.9% for our collar detectors and 0.8% for the eV crystal at 662 keV were obtained. The latter compares favourably to the best existing energy resolution results from pixel detectors.
Explore related subjects
Keep this discovery
Y. Ramachers, D. Y. Stewart. 2007-09-21. Energy resolution improvement in room-temperature CZT detectors. https://doi.org/10.1088/1748-0221/2/12/p12003
Cite the original work for its findings. Save a collection to share your selection of sources.