Conduction Effect of Thermal Radiation in a Metal Shield Pipe in a Cryostat for a Cryogenic Interferometric Gravitational Wave Detector
Abstract
A large heat load caused by thermal radiation through a metal shield pipe was observed in a cooling test of a cryostat for a prototype of a cryogenic interferometric gravitational wave detector. The heat load was approximately 1000 times larger than the value calculated by the Stefan-Boltzmann law. We studied this phenomenon by simulation and experiment and found that it was caused by the conduction of thermal radiation in a metal shield pipe.
Explore related subjects
Keep this discovery
Takayuki Tomaru, Masao Tokunari, Kazuaki Kuroda, Takashi Uchiyama, Akira Okutomi, Masatake Ohashi, Hiroyuki Kirihara, Nobuhiro Kimura, Yoshio Saito, Nobuaki Sato, Takakazu Shintomi, Toshikazu Suzuki, Tomiyoshi Haruyama, Shinji Miyoki, Kazuhiro Yamamoto, Akira Yamamoto. 2007-11-06. Conduction Effect of Thermal Radiation in a Metal Shield Pipe in a Cryostat for a Cryogenic Interferometric Gravitational Wave Detector. https://doi.org/10.1143/jjap.47.1771