Detector for imaging of explosions: present status and future prospects with higher energy X-rays
Abstract
The detector for imaging of explosions (DIMEX) is in operation at the synchrotron radiation (SR) beam-line at VEPP-3 electron ring at Budker INP since 2002. DIMEX is based on one-coordinate gas ionization chamber filled with Xe-CO2(3:1) mixture at 7atm, and active Frisch-grid made of Gas Electron Multiplier (GEM). The detector has spatial resolution of ~0.2mm and dynamic range of ~100 that allows to realize the precision of signal measurement at a percent level. The frame rate can be tuned up to 8 MHz (125 ns per image) and up to 32 images can be stored in one shot. At present DIMEX is used with the X-ray beam from 2T wiggler that has ~20 keV average energy. Future possibility to install similar detector at the SR beam-line at VEPP-4 electron ring is discussed.
Explore related subjects
Keep this discovery
V. M. Aulchenko, O. V. Evdokov, L. I. Shekhtman, K. A. Ten, B. P. Tolochko, I. L. Zhogin, V. V. Zhulanov. 2008-01-15. Detector for imaging of explosions: present status and future prospects with higher energy X-rays. https://doi.org/10.1088/1748-0221/3/05/p05005