High-precision efficiency calibration of a high-purity co-axial germanium detector
Abstract
A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and gamma-ray source measurements have been compared to Monte-Carlo simulations to adjust the dimensions of a detector model. For this purpose, standard calibration sources and short-lived on-line sources have been used. The resulting efficiency calibration reaches the precision needed e.g. for branching ratio measurements of super-allowed beta decays for tests of the weak-interaction standard model.
Explore related subjects
Keep this discovery
B. Blank, J. Souin, P. Ascher, L. Audirac, G. Canchel, M. Gerbaux, S. Grevy, J. Giovinazzo, H. Guerin, T. Kurtukian Nieto, I. Matea, H. Bouzomita, P. Delahaye, G. F. Grinyer, J. C. Thomas. 2014-04-04. High-precision efficiency calibration of a high-purity co-axial germanium detector. https://doi.org/10.1016/j.nima.2014.12.071