arXiv · 1501.05608
Ion behavior and interelectrode breakdown voltage of a drift tube
Abstract
We experimentally studied ion behavior and interelectrode breakdown voltage. The ion behavior of a drift tube directly influences the detection of ion intensity, and then influences the detection sensitivity of a system. Interelectrode voltage and pressure directly influence the ion behavior. Gas discharge between electrodes influences the adjustments required for interelectrode voltage. The experimental results show: ion intensity increases exponentially with the increment of voltage between drift electrodes; ion intensity decreases exponentially as pressure increases; with the increment of pressure, the breakdown voltage at first decreases, and then increases; ion injection has a significant influence on breakdown voltage, and this influence depends on the pressure and shapes of the electrodes. We explain the results above through assumptions and by mathematical methods.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Hao Geng, Zhongjun Zhao, Yixiang Duan. 2015-01-12. Ion behavior and interelectrode breakdown voltage of a drift tube. https://doi.org/10.1088/1674-1137%2F39%2F5%2F057003
Cite the original work for its findings. Save a collection to share your selection of sources.