A new method of energy calibration of position-sensitive silicon detector
Abstract
An improved method of energy calibration of position-sensitive silicon detector is presented. Instead of the parabolic function used in traditional method, a new function describing the relation of position and energy is introduced and achieves better energy resolution. For the 8.088 MeV alpha decay of 213Rn calibrated by this new method, the energy resolution is determined to be about 87 keV (FWHM), which is better than the result of the traditional method, 104 keV (FWHM). In addition, different functions can be tried in the new method, which makes the calibration of various detectors with different performances possible.