Spatial resolution measurement of Triple-GEM detector and diffraction imaging test at synchrotron radiation
Abstract
A triple-GEM detector with two-dimensional readout is developed. The detector provides high position resolution for powder diffraction experiments at synchrotron radiation. Spatial resolution of the detector is measured in the lab using a 55Fe X-ray source. A resolution of about 110 um FWHM is achieved. The energy resolution is better than 27% for 5.9 keV X-rays. The detector's validity under illumination of photons in particular energy range is verified using a Cu X-ray tube. Imaging of the head of a wire stripper with X-ray tube demonstrates its imaging ability. A diffraction imaging experiment using the sample of powder SiO2 is successfully carried out at 1W2B laboratory of Beijing Synchrotron Radiation Facility (BSRF). Different diffraction rings are clearly seen under various X-ray energies.
Explore related subjects
Keep this discovery
Y. L. Zhang, H. R. Qi, Z. W. Wen, H. Y. Wang, Q. Ouyang, Y. B. Chen, J. Zhang, B. T. Hu. 2017-02-10. Spatial resolution measurement of Triple-GEM detector and diffraction imaging test at synchrotron radiation. https://doi.org/10.1088/1748-0221/12/04/p04015
Cite the original work for its findings. Save a collection to share your selection of sources.