High-fidelity, high-isotropic resolution diffusion imaging through gSlider acquisition with B1+ & T1 corrections and integrated ΔB0/Rx shim array
Abstract
Purpose: B1+ and T1 corrections and dynamic multi-coil shimming approaches were proposed to improve the fidelity of high isotropic resolution Generalized slice dithered enhanced resolution (gSlider) diffusion imaging. Methods: An extended reconstruction incorporating B1+ inhomogeneity and T1 recovery information was developed to mitigate slab-boundary artifacts in short-TR gSlider acquisitions. Slab-by-slab dynamic B0 shimming using a multi-coil integrated ΔB0/Rx shim-array, and high in-plane acceleration (Rinplane=4) achieved with virtual-coil GRAPPA were also incorporated into a 1 mm isotropic resolution gSlider acquisition/reconstruction framework to achieve an 8-11 fold reduction in geometric distortion compared to single-shot EPI. Results: The slab-boundary artifacts were alleviated by the proposed B1+ and T1 corrections compared to the standard gSlider reconstruction pipeline for short-TR acquisitions. Dynamic shimming provided >50% reduction in geometric distortion compared to conventional global 2nd order shimming. 1 mm isotropic resolution diffusion data show that the typically problematic temporal and frontal lobes of the brain can be imaged with high geometric fidelity using dynamic shimming. Conclusions: The proposed B1+ and T1 corrections and local-field control substantially improved the fidelity of high isotropic resolution diffusion imaging, with reduced slab-boundary artifacts and geometric distortion compared to conventional gSlider acquisition and reconstruction. This enabled high-fidelity whole-brain 1 mm isotropic diffusion imaging with 64 diffusion-directions in 20 minutes using a 3T clinical scanner.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Congyu Liao, Jason Stockmann, Qiyuan Tian, Berkin Bilgic, Nicolas S. Arango, Mary Kate Manhard, William A. Grissom, Lawrence L. Wald, Kawin Setsompop. 2019-03-26. High-fidelity, high-isotropic resolution diffusion imaging through gSlider acquisition with B1+ & T1 corrections and integrated ΔB0/Rx shim array. https://doi.org/10.1002/mrm.27899
Cite the original work for its findings. Save a collection to share your selection of sources.