arXiv · 1907.03898
Parametrically Amplified Low-Power MEMS Capacitive Humidity Sensor
Abstract
We present the design, fabrication, and response of a polymer-based Laterally Amplified Chemo-Mechanical (LACM) humidity sensor based on mechanical leveraging and parametric amplification. The device consists of a sense cantilever asymmetrically patterned with a polymer and flanked by two stationary electrodes on the sides. When exposed to a humidity change, the polymer swells after absorbing the analyte and causes the central cantilever to bend laterally towards one side, causing a change in the measured capacitance. The device features an intrinsic gain due to parametric amplification resulting in an enhanced signal-to-noise ratio (SNR). 11-fold magnification in sensor response was observed via voltage biasing of the side electrodes without the use of conventional electronic amplifiers. The sensor showed a repeatable and recoverable capacitance change of 11% when exposed to a change in relative humidity from 25-85%. The dynamic characterization of the device also revealed a response time ~1s and demonstrated a competitive response with respect to a commercially available reference chip.
Explore related subjects
Keep this discovery
Rugved Likhite, Aishwaryadev Banerjee, Apratim Majumder, Hanseup Kim and, Carlos H. Mastrangelo. 2019-07-08. Parametrically Amplified Low-Power MEMS Capacitive Humidity Sensor. https://doi.org/10.3390/s19183954
Cite the original work for its findings. Save a collection to share your selection of sources.