arXiv · 1909.11431
Prototype design of singles processing unit for the small animal PET
Abstract
Position Emission Tomography (PET) is an advanced clinical diagnostic imaging technique for nuclear medicine. Small animal PET is increasingly uesd for studying the animal model of disease, new drugs and new therapies. A prototype of Singles Processing Unit (SPU) for a small animal PET system was designed to obtain the time, energy, and position information. The energy and position is actually calculated through high precison charge measurement, which is based on amplification, shaping, A/D conversion and area calculation in digital signal processing domian. Analysis and simulations were also conducted to optimize the key parameters in system design. Initial tests indicate that the charge and time precision is better than 0.3% FWHM and 350 ps FWHM respectively, while the position resolution is better than 0.35% FWHM. Commination tests of the SPU prototype with the PET detector indicate that the system time precision is better than 2.5 ns, while the flood map and energy spectra concored well with the expected.
Explore related subjects
Keep this discovery
Peipei Deng, Lei Zhao, Jiaming Lu, Bowen Li, Ruoshi Dong, Shubin Liu, Qi An. 2019-09-25. Prototype design of singles processing unit for the small animal PET. https://arxiv.org/abs/1909.11431
Cite the original work for its findings. Save a collection to share your selection of sources.