arXiv · 2110.01531
Measuring the Quantum Efficiency of X-Ray Hybrid CMOS Detectors
Abstract
Next-generation X-ray observatories, such as the Lynx X-ray Observatory Mission Concept, will require detectors with high quantum efficiency (QE) across the soft X-ray band to observe the faint objects that drive their mission science cases. Hybrid CMOS Detectors (HCDs), a form of active-pixel sensor, are promising candidates for use on these missions because of their faster read-out, lower power consumption, and greater radiation hardness than detectors used in the current generation of X-ray telescopes. In this work, we present QE measurements of a Teledyne H2RG HCD. These measurements were performed using a gas-flow proportional counter as a reference detector to measure the absolute flux incident on the HCD. We find an effective QE of $95.0 \pm 1.1\%$ at the Mn K$\alpha$/K$\beta$ lines (at 5.9 and 6.5 keV), $98.5 \pm 1.8\%$ at the Al K$\alpha$ line (1.5 keV), and $85.0 \pm 2.8\%$ at the O K$\alpha$ line (0.52 keV).
Explore related subjects
Keep this discovery
Joseph M. Colosimo, Abraham D. Falcone, Mitchell Wages, Samuel V. Hull, David N. Burrows, Mitchell Range, Fredric Hancock, Cole R. Armstrong, Gooderham McCormick, Daniel M. LaRocca. 2021-10-04. Measuring the Quantum Efficiency of X-Ray Hybrid CMOS Detectors. https://doi.org/10.1117/12.2594040
Cite the original work for its findings. Save a collection to share your selection of sources.