External beam irradiation angle measurement using Cerenkov emission

Abstract

Due to its angular dependency, Cerenkov light has long been considered a contamination signal in plastic scintillating dosimeters. In this study, we propose a novel approach designed to take advantage of this angular dependency to perform a direct measurement of an external beam radiation angle of incidence. A Cerenkov probe composed of a 10-mm long filtered sensitive volume of clear PMMA optical fibre was built. Both filtered and raw Cerenkov signals from the transport fibre were collected through a single 1-mm diameter transport fibre. An independent PSD composed of 10-mm BCF12 scintillating fibre was also used for simultaneous dose measurements. A first series of measurements aimed at validating the ability to account for the Cerenkov electron energy spectrum dependency by simultaneously measuring the deposited dose. A cylindrical phantom was then used to obtain an angular calibration curve for fixed dose irradiations and perform incident angle measurements using electron and photon beams. The beam nominal energy was found to have a significant impact on the shapes of the angular calibration curves obtained for irradiation angle measurements. This can be linked to the electron energy spectrum dependency of the Cerenkov emission cone. Irradiation angle measurements exhibit an absolute mean error of 1.86{\deg} and 1.02{\deg} at 6 and 18 MV, respectively. Similar results were obtained with electron beams and the absolute mean error reaches 1.97, 1.66, 1.45 and 0.95 degrees at 9, 12, 16 and 20 MeV, respectively. Reducing the numerical aperture of the Cerenkov probe leads to an increased angular dependency for the lowest energy while no major changes were observed at higher energy. This allowed irradiation angle measurements at 6 MeV with a mean absolute error of 4.82{\deg}. The detector offers promising perspectives for external beam radiotherapy and brachytherapy applications.

Explore related subjects

Keep this discovery

BibTeXRIS

Emilie Jean, Simon Lambert-Girard, Francois Therriault-Proulx, Luc Beaulieu. 2021-11-22. External beam irradiation angle measurement using Cerenkov emission. https://arxiv.org/abs/2111.11304

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph