SearcharxivSearch

arXiv · 2206.02375

Performance of TPC detector prototype integrated with UV laser tracks for the circular collider

Abstract

Several new experimental concepts in high-energy particle physics have been proposed in recent years. The physical goals include precisely measuring the properties of particles such as Higgs, Z and W, and even looking for signs of new physics at future colliders. To meet the evolving requirements for particle track detector, Time Projection Chamber(TPC) detector prototype integrated with a UV laser track system was developed for the main track detector at Circular Electron Positron Collider(CEPC). This prototype consists of 6 horizontal laser tracks around TPC detector chamber, a fast electronics readout of 1280 channels, a GEM detector with $200\times 200\,mm^2$ active area, and the DAQ system. The hit resolution, dE/dx resolution and drift velocity were studied by measuring and analyzing using the TPC prototype and UV laser tracks. The dE/dx resolution of the prototype was measured to be $(8.9\pm0.4)\,\%$. Extrapolating this to CEPC TPC with 220 layers and longer track, the resolution was estimated to be $(3.36\pm0.26)\,\%$. All results indicated that the TPC detector prototype integrated with UV laser tracks can work well.

Explore related subjects

Keep this discovery

BibTeXRIS

Z. Y. Yuan, H. R. Qi, Y. Chang, L. W. Yu, Y. M. Cai, H. Y. Zhang, J. Zhang, Q. Ouyang, Y. L. Li, Z. Deng, H. Gong. 2022-06-06. Performance of TPC detector prototype integrated with UV laser tracks for the circular collider. https://doi.org/10.1016/j.nima.2022.167241

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det