A Retrofittable Photoelectron Gun Proposal for Improved Low Voltage Imaging
Abstract
Low voltage scanning electron microscopy is a powerful tool for examining surface features and imaging beam sensitive materials. Improving resolution during low voltage imaging is then an important area of development. Decreasing the effect of chromatic aberration is one solution to improving the resolution and can be achieved by reducing the energy spread of the electron source. Our approach involves retrofitting a light source onto a thermionic Lanthanum Hexaboride (LaB6) electron gun as a cost-effective low energy spread photoelectron emitter. The energy spread of the emitter's photoelectrons is predicted to be between 0.11-0.38eV depending on the photon energy of the UV light source. Proof of principle images have been recorded using this retrofitted photoelectron gun and an analysis of its performance is presented.
Explore related subjects
Keep this discovery
Frances Quigley, Clive Downing, Cormac McGuinness, Lewys Jones. 2022-12-06. A Retrofittable Photoelectron Gun Proposal for Improved Low Voltage Imaging. https://arxiv.org/abs/2212.02964
Cite the original work for its findings. Save a collection to share your selection of sources.