Towards 3D Magnetic Force Microscopy
Abstract
Magnetic force microscopy (MFM) is long established as a powerful tool for probing the local manifestation of magnetic nanostructures across a range of temperatures and applied stimuli. A major drawback of the technique, however, is that the detection of stray fields emanating from a samples surface rely on a uniaxial vertical cantilever oscillation, and thus are only sensitive to vertically oriented stray field components. The last two decades have shown an ever-increasing literature fascination for exotic topological windings where particular attention to in-plane magnetic moment rotation is highly valuable when identifying and understanding such systems. Here we present a new method of detecting in-plane magnetic stray field components, by utilizing a home made split-electrode excitation piezo that allows the simultaneous excitation of a cantilever at its fundamental flexural and torsional modes. This allows for the joint acquisition of traditional vertical mode (V-MFM) images and a lateral MFM (L-MFM) where the tip-cantilever system is only sensitive to stray fields acting perpendicular to the torsional axis of the cantilever.