Correcting impedance measurements for background parasitics to characterize circuit components in cryogenic environments
Abstract
Predictable circuit response is a critical prerequisite for accurate electronic measurements. We describe a powerful, yet straightforward, experimental method and analysis model that utilizes an affordable LCR meter in conjunction with an in situ parasitic impedance background correction procedure to measure the temperature-dependent impedance (magnitude and phase) of individual passive circuit elements mounted in a cryostat. We show how the model unambiguously identified a 20x drop in capacitance for 22 microF 5XR multilayer ceramic capacitors cooled from 300 K to 360 mK in an environment with parasitic capacitance of order 300 pF. The same experimental procedure, based on a simple two-wire measurement, was also used to successfully measure 10 pF and 22 pF thin-film capacitors and 100 MOhm thick-film resistors. The results showed that the resistor values increased by up to an order of magnitude when the devices were cooled from 300 K to 360 mK. Most importantly, the simple data acquisition method and robust analysis model were shown to effectively extend the accuracy of a simple benchtop LCR meter beyond its manufacturer-guaranteed values for a wide range of measurement frequencies.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Riley J. Carpenter, Jadyn Anczarski, Ivar Rydstrom, Aviv Simchony, Zoe J. Smith, Noah A. Kurinsky, Betty A. Young. 2025-06-13. Correcting impedance measurements for background parasitics to characterize circuit components in cryogenic environments. https://arxiv.org/abs/2506.12268
Cite the original work for its findings. Save a collection to share your selection of sources.