Characterization of a Large-Area HRPPD MCP-PMT Coupled to LYSO for TOF-PET

Abstract

Large-area microchannel-plate photomultipliers (MCP-PMTs) such as the High-Rate Picosecond Photon Detector (HRPPD) are attractive for time-of-flight positron emission tomography (TOF-PET). They offer excellent single-photon time resolution, low dark count rate, and large monolithic active areas. However, the coincidence time resolution (CTR) reported so far with PET scintillators falls short of expectations based upon their single-photon time resolution. We characterized a 104x104 mm$^2$ HRPPD coupled to LYSO crystals and read out through a custom high-frequency amplifier. With a 2x2x3 mm$^3$ LYSO crystal we measured a CTR of 107 ps full width at half maximum (FWHM) against a silicon photomultiplier (SiPM)-based reference detector. After subtracting the calibrated reference detector's contribution of 42 ps, the single HRPPD detector was found to contribute 98 ps, corresponding to 139 ps for two identical HRPPD detectors in coincidence. With clinical 3x3x20 mm$^3$ crystals the CTR against the reference SiPM detector was 160 ps, corresponding to 218 ps for two identical HRPPD detectors. Timing and gain were uniform across output pads within measurement precision. Summing the four pads under a 3x3x10 mm$^3$ LYSO crystal, the energy resolution for 511 keV gamma ray was 14.6% FWHM, and the 1275 keV line of the $^{22}$Na source was within 1% of its proportional position relative to the 511 keV peak. Flood maps resolved the individual 3 mm elements of a 4x4 array despite the 25 mm pad pitch. To the best of our knowledge the observed timing resolution improves substantially on what has been reported for large-area MCP-PMTs coupled to PET scintillators, and the crystal array identification has not been demonstrated before on such a device. These results indicate that this device class deserves renewed consideration as a photodetector technology for high-performance TOF-PET systems.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Yingjie Wang, Mark A. Popecki, Heejong Kim, Alexey V. Lyashenko, Stefan Cwik, Mythra Varun Nemallapudi, Jeffrey S. Souris, Michael J. Minot, Chien-Min Kao, Chin-Tu Chen. 2026-11-11. Characterization of a Large-Area HRPPD MCP-PMT Coupled to LYSO for TOF-PET. https://searcharxiv.com/abs/2611.15820

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Searcharxiv by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related papers

Improved Pion-Kaon Identification in Heavy-Ion Collisions with a Two-Dimensional Transformation

Accurate identification of charged pions and kaons is essential for precision measurements in relativistic heavy-ion collisions, but becomes increasingly challenging at intermediate and high transverse momentum because of the increasing overlap of the TPC and TOF particle identification responses. In this work, we quantitatively benchmark a two-dimensional shift-and-rotation PID procedure that combines the $nσ_π$ and mass-squared $(m^{2})$ information. The benchmark is performed using Au+Au collision events generated with the AMPT model, where the detector response is modeled using $p_{T}$-dependent PID parameters extracted from STAR data. Since the true particle identities and input yields are known in AMPT but are not used during the reconstruction, the procedure can be evaluated through a direct closure test that is not possible with experimental data alone. The two-dimensional procedure is compared with a reference one-dimensional method consisting of an $|nσ_π|<$ 3 preselection followed by a Gaussian fit to the $m^{2}$ distribution. While both methods perform well when pion and kaon PID responses are well separated, the two-dimensional procedure provides improved yield closure as the PID distributions increasingly overlap at high $p_{T}$. The reconstructed-to-input yield ratio remains within 2\% of unity up to $p_T =$ 3 GeV/$c$, whereas the reference one-dimensional procedure shows a larger deviation. Reaction-plane $v_{2}$ closure tests show that both the reference one-dimensional and two-dimensional PID procedures reproduce the AMPT input $v_{2}$ within the statistical uncertainties over the studied $p_{T}$ range. These results demonstrate that the two-dimensional procedure substantially improves the high-$p_{T}$ yield reconstruction without introducing a detectable bias in the extracted $v_{2}$ under the STAR-like detector response considered here.

physics.ins-det↗

Cryogenic Mechanical Loss of GaAs/AlGaAs Crystalline Coatings

Optical coatings with minimal thermal noise are essential for high-precision, laser-based metrology experiments. Substrate-transferred crystalline GaAs/Al$_{0.92}$Ga$_{0.08}$As coatings have the lowest thermal noise for high-reflectivity mirror coatings with $λ=900$ nm -- 12 $μ$m. Time-frequency stabilization experiments have demonstrated that GaAs/AlGaAs mirror coatings have a lower noise floor than amorphous mirror coatings. These mirrors are now being developed at sizes $\ge 300$ mm for gravitational-wave detectors. In order to understand the coating's fundamental loss mechanisms, which cause thermal noise, we have performed the first measurement of the mechanical loss at cryogenic temperatures using a cryogenic, multimodal, gentle nodal suspension system, which operates from 12 K to room temperature. This initial measurement used a silicon substrate whose asymmetry produced excess friction with the nodal support. Accounting for this friction and Akhiezer loss in the substrate and coating, the mechanical loss agrees well with the thermal noise observed in experiments using cryogenic fixed-cavities with GaAs/AlGaAs mirrors.

physics.ins-det↗

Rotating Capacitive System for Measurement of a Third Order Gravity Gradient Based on a Single-Axis Mechanical Oscillator. Part I, Theory

This paper is a feasibility study on a single-axis mechanical oscillator, mounted on a miniature circular rotating platform, that is proposed to be used for measuring a quantity proportional to the second vertical derivative of gravitational acceleration projected onto the oscillator's sensitivity axis. The latter is aligned with the axis of rotation. Part I establishes theoretical feasibility only. Design constraints, quantitative sensitivity, error budgets, and mobile platform performance will be addressed in Part II.

physics.ins-det↗