arXiv · 1602.01702
Design and Performance of a Lead Fluoride Detector as a Luminosity Monitor
Abstract
Precise luminosity measurements for the OLYMPUS two-photon exchange experiment at DESY were performed by counting scattering events with alternating beams of electrons and positrons incident on atomic electrons in a gaseous hydrogen target. Final products of M{\o}ller, Bhabha, and pair annihilation interactions were observed using a pair of lead fluoride Cherenkov calorimeters with custom housings and electronics, adapted from a system used by the A4 parity violation experiment at MAMI. This paper describes the design, calibration, and operation of these detectors. An explanation of the Monte Carlo methods used to simulate the physical processes involved both at the scattering vertices and in the detector apparatus is also included.
Explore related subjects
Keep this discovery
Roberto Pérez Benito, Dmitry Khaneft, Colton O'Connor, Luigi Capozza, Jürgen Diefenbach, Boris Gläser, Yue Ma, Frank Maas, David Rodríguez Piñeiro. 2016-02-04. Design and Performance of a Lead Fluoride Detector as a Luminosity Monitor. https://doi.org/10.1016/j.nima.2016.04.071
Cite the original work for its findings. Save a collection to share your selection of sources.