arXiv · 2001.04854
Hyperfine spectroscopy in a quantum-limited spectrometer
Abstract
We report measurements of electron spin echo envelope modulation (ESEEM) performed at millikelvin temperatures in a custom-built high-sensitivity spectrometer based on superconducting micro-resonators. The high quality factor and small mode volume (down to 0.2pL) of the resonator allow to probe a small number of spins, down to $5\cdot 10^2$. We measure 2-pulse ESEEM on two systems: erbium ions coupled to $^{183}$W nuclei in a natural-abundance $\text{CaWO}_4$ crystal, and bismuth donors coupled to residual $^{29}$Si nuclei in a silicon substrate that was isotopically enriched in the $^{28}$Si isotope. We also measure 3- and 5-pulse ESEEM for the bismuth donors in silicon. Quantitative agreement is obtained for both the hyperfine coupling strength of proximal nuclei, and the nuclear spin concentration.
Explore related subjects
Keep this discovery
S. Probst, G. L. Zhang, M. Rančić, V. Ranjan, M. Le Dantec, Z. Zhong, B. Albanese, A. Doll, R. B. Liu, J. J. L. Morton, T. Chanelière, P. Goldner, D. Vion, D. Esteve, P. Bertet. 2020-01-14. Hyperfine spectroscopy in a quantum-limited spectrometer. https://doi.org/10.5194/mr-1-315-2020
Cite the original work for its findings. Save a collection to share your selection of sources.