Frequency Shift Algorithm: Design of a Baseband Phase Locked Loop for Frequency-Domain Multiplexing Readout of X-ray Transition-Edge Sensor Microcalorimeters

Abstract

The Transition-Edge Sensor (TES) is an extremely sensitive device which is used to measure the energy of individual X-ray photons. For astronomical spectrometry applications, SRON develops a Frequency Domain Multiplexing (FDM) read-out system for kilopixel arrays of such TESs. Each TES is voltage biased at a specific frequency in the range 1 to 5 MHz. Isolation between the individual pixels is obtained through very narrow-band (high-Q) lithographic LC resonators. To prevent energy resolution degradation due to intermodulation line noise, the bias frequencies are distributed on a regular grid. The requirements on the accuracy of the LC resonance frequency are very high. The deviation of the resonance frequencies due to production tolerances is significant with respect to the bandwidth, and a controller is necessary to compensate for the LC series impedance. We present two such controllers: a simple orthogonal proportional-integrating (PI) controller and a more complex impedance estimator. Both controllers operate in baseband and try to make the TES current in-phase with the bias voltage, effectively operating as phase-locked loops (PLL). They allow off-LC-resonance operation of the TES pixels, while preserving TES thermal response and energy resolution. Extensive experimental results -- published in a companion paper recently -- with the proposed methods, show that these controllers allow the preservation of single pixel energy resolution in multiplexed operation.

Explore related subjects

Keep this discovery

BibTeXRIS

Paul van der Hulst, Jan van der Kuur, Ad Nieuwenhuizen, Davide Vaccaro, Hiroki Akamatsu, Patrick van Winden, Bert-Joost van Leeuwen, Jan-Willem den Herder. 2021-07-02. Frequency Shift Algorithm: Design of a Baseband Phase Locked Loop for Frequency-Domain Multiplexing Readout of X-ray Transition-Edge Sensor Microcalorimeters. https://doi.org/10.1063/5.0044968

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det