arXiv · 2309.15011
Characterization of LAPPD timing at CERN PS testbeam
Abstract
Large Area Picosecond PhotoDetectors (LAPPDs) are photosensors based on microchannel plate technology with about 400 cm$^2$ sensitive area. The external readout plane of a capacitively coupled LAPPD can be segmented into pads providing a spatial resolution down to 1 mm scale. The LAPPD signals have about 0.5 ns risetime followed by a slightly longer falltime and their amplitude reaches a few dozens of mV per single photoelectron. In this article, we report on the measurement of the time resolution of an LAPPD prototype in a test beam exercise at CERN PS. Most of the previous measurements of LAPPD time resolution had been performed with laser sources. In this article we report time resolution measurements obtained through the detection of Cherenkov radiation emitted by high energy hadrons. Our approach has been demonstrated capable of measuring time resolutions as fine as 25-30 ps. The available prototype had performance limitations, which prevented us from applying the optimal high voltage setting. The measured time resolution for single photoelectrons is about 80 ps r.m.s.
Explore related subjects
Keep this discovery
Deb Sankar Bhattacharya, Andrea Bressan, Chandradoy Chatterjee, Silvia Dalla Torre, Mauro Gregori, Alexander Kiselev, Stefano Levorato, Anna Martin, Saverio Minutoli, Mikhail Osipenko, Richa Rai, Marco Ripani, Fulvio Tessarotto, Triloki Triloki. 2023-09-26. Characterization of LAPPD timing at CERN PS testbeam. https://arxiv.org/abs/2309.15011
Cite the original work for its findings. Save a collection to share your selection of sources.