arXiv · 2411.00501
Design and Implementation of a Compact Analog Constant Fraction Discriminator for High-Resolution Timing in Gamma-Ray Spectroscopy
Abstract
This work presents a custom analog front-end card designed for the read-out of PMTs coupled to lanthanum bromide scintillators. It integrates 16 discrete analog constant fraction discriminators (CFDs) on a compact 12x10 cm board, providing precise timing information for nuclear lifetime measurements. The design emphasizes cost-effectiveness, utilizing off-the-shelf discrete components, as well as compactness, achieved by using miniature coaxial connectors and cables as delay elements. The focus of this paper lies on the unusual and extremely minimalist analog shaper/discriminator design which is devised without operational amplifiers, making use of only a handful of RF transistors and LVDS receivers in place of comparators.
Explore related subjects
Keep this discovery
Michael Wiebusch. 2024-11-01. Design and Implementation of a Compact Analog Constant Fraction Discriminator for High-Resolution Timing in Gamma-Ray Spectroscopy. https://arxiv.org/abs/2411.00501
Cite the original work for its findings. Save a collection to share your selection of sources.