arXiv · 2608.26542
Measuring and Modelling Lag in Amorphous Silicon Flat-Panel X-ray Detectors
Abstract
Detector lag, also referred to as afterglow, is a source of image degradation in flat-panel x-ray detectors, producing temporal artefacts that reduce image quality and quantitative accuracy. In this work, we present a robust and repeatable experimental framework for characterising long-term detector lag (from minutes to hours) under controlled step-up and step-down exposure transitions of particular relevance to tomography. The measured transition curves from an amorphous silicon detector with a CsI:Ti scintillator exhibit unexpected behaviour, with the detector response temporarily overshooting its final equilibrium intensity following step-up transitions, rather than exhibiting the gradual monotonic rise and decay typically expected. These curves were well fitted by multi-exponential functions, providing initial estimates for the depth of the charge traps in the scintillator. These parameters were incorporated into a proposed multi-trap rate equation model that reproduces the overall observed behaviour of both step-up and step-down transitions, including the overshoot that is not captured by previous models. Although discrepancies remain between the model and observation, this work establishes a reproducible methodology for detector lag characterisation and presents an improved physical model that offers greater insight into the mechanisms governing detector lag.
Explore related subjects
Keep this discovery
Yiyue Huang, Benjamin Young, Andrew Kingston, Adrian Sheppard. 2026-08-27. Measuring and Modelling Lag in Amorphous Silicon Flat-Panel X-ray Detectors. https://arxiv.org/abs/2608.26542
Cite the original work for its findings. Save a collection to share your selection of sources.