Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance

Abstract

A liquid isooctane (C$_{8}$H$_{18}$) filled ionization linear array for radiotherapy quality assurance has been designed, built and tested. The detector consists of 128 pixels, each of them with an area of 1.7 mm $\times$ 1.7 mm and a gap of 0.5 mm. The small pixel size makes the detector ideal for high gradient beam profiles like those present in Intensity Modulated Radiation Therapy (IMRT) and radiosurgery. As read-out electronics we use the X-Ray Data Acquisition System (XDAS) with the Xchip developed by the CCLRC. Studies concerning the collection efficiency dependence on the polarization voltage and on the dose rate have been made in order to optimize the device operation. In the first tests we have studied dose rate and energy dependences, and signal reproducibility. Dose rate dependence was found lower than 2.5 % up to 5 Gy min$^{-1}$, and energy dependence lower than 2.1 % up to 20 cm depth in solid water. Output factors and penumbras for several rectangular fields have been measured with the linear array and were compared with the results obtained with a 0.125 cm$^{3}$ air ionization chamber and radiographic film, respectively. Finally, we have acquired profiles for an IMRT field and for a virtual wedge. These profiles have also been compared with radiographic film measurements. All the comparisons show a good correspondence. Signal reproducibility was within a 2% during the test period (around three months). The device has proved its capability to verify on-line therapy beams with good spatial resolution and signal to noise ratio.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

J. Pardo, L. Franco, F. Gomez, A. Iglesias, A. Pazos, J. Pena, R. Lobato, J. Mosquera, M. Pombar, J. Sendon. 2004-12-01. Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance. https://doi.org/10.1088/0031-9155%2F50%2F8%2F006

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Light-sheet microscopy in cancer pathology: current applications, clinical challenges and future perspectives

Light-sheet fluorescence microscopy (LSFM) has emerged as a powerful research technology for three-dimensional (3D) visualization of biological tissues and is increasingly applied to human cancer specimens, including optically cleared biopsies and surgical resection samples. In this Perspective article, we review current applications of LSFM in oncology, and discuss how recent developments in microscopy, tissue preparation, labelling, and computational analysis are expanding the possibilities of 3D cancer pathology. Although LSFM has demonstrated the feasibility of large-volume imaging in selected human specimens, its application remains predominantly within research and translational settings and is not currently part of routine clinical pathology. We therefore distinguish demonstrated applications from potential future uses. We discuss how LSFM could eventually contribute to selected diagnostic and surgical workflows by providing 3D information on tumour architecture, spatial heterogeneity, metastatic dissemination, and surgical margins. These applications remain prospective and will require substantial technical, clinical, regulatory, computational, and workflow validation before clinical implementation can be considered. Ultimately, LSFM may complement rather than replace conventional histopathology by extending pathological assessment from two-dimensional tissue sections towards 3D spatial analysis.

physics.med-ph↗

Quantitative Ultrasound Perfusion Biomarker: Cross-Modality Validation with Laser Speckle Imaging

Rationale and Objectives: Tooth extraction can cause alveolar bone resorption and variable healing despite socket preservation. Current perfusion assessment relies largely on visual evaluation, limiting objective monitoring of socket healing. We developed and validated a quantitative color flow ultrasound framework to reduce sensitivity to imaging settings and motion artifacts. Materials and Methods: Color flow ultrasound data were acquired longitudinally in 30 patients undergoing tooth extraction and grafting. Velocity- and power-weighted color pixel density (CPDvel and CPDpwr) were derived from complex-valued color flow (TRI) data at 0-1.5 mm beneath the epithelial surface. Gain robustness was assessed using a Quantitative Imaging Biomarker Alliance (QIBA) flow phantom. Concurrent laser speckle contrast imaging (LASCA) provided in vivo comparison. Results: Phantom velocity estimates remained stable across color gain settings of 20-100 a.u. at flow rates of 2, 4, 6, and 8 mL/s. Perfusion metrics decreased over approximately 2-30 weeks after extraction. Nonlinear fits yielded R2 values of 0.74 for CPDvel, 0.45 for CPDpwr, and 0.66 for LASCA. Exponential half-decay times were 60.3, 92.4, and 48.3 days, respectively, relative to the first post-extraction measurement. LASCA correlated linearly with ultrasound-derived velocity (R2 = 0.68) and power (R2 = 0.74; both p < 0.05). Conclusions: The framework supports noninvasive, quantitative monitoring of superficial perfusion during socket healing with reduced sensitivity to color gain. Direct analysis of TRI data provides depth-specific perfusion metrics and may improve consistency across imaging sessions. Further validation is needed to determine its ability to identify compromised healing and inform clinical decisions.

physics.med-ph↗

LesionStudio: A Patient-Informed Platform for Task-Specific Optimization of Emission Tomography Reconstructions

Purpose: For task-specific PET and SPECT reconstruction optimization, we developed LesionStudio, an open-source platform incorporating patient background heterogeneity, realistic lesion morphology and texture, and lesion-background interactions beyond conventional phantoms. Methods: LesionStudio, a Python framework built on PyTomography, inserts lesions into PET/SPECT data using modality-specific forward projection and stochastic background ablation. Clinical realism was evaluated by physician review of 30 inserted lesions in eight DLBCL patients from a 1,579-lesion FDG lymphoma library. Quantification and detectability were assessed in FDG PET using spherical and patient-derived lesions varying in size, LBR, and site, reconstructed with OSEM and BSREM. Results: Inserted lesions were visually indistinguishable from true disease (AUC = 0.49). OSEM improved the patient-derived quantification limit from ~3.7 mL at iteration 4 to ~2.0 mL at iteration 10, but with increased noise and edge artifacts. Weakly regularized BSREM (beta = 5) achieved a similar ~2.0 mL limit with stability, whereas stronger regularization increased under-recovery. Detectability was contrast-dependent: no lesions were detected at LBR 2:1, whereas detectable volume ranged from 0.70 to 3.63 mL at LBR 4:1 and decreased to 0.70 mL at LBR 6:1-8:1 for both OSEM and weakly regularized BSREM. Conclusion: Reconstruction settings can shift SUV metrics and detectability enough to affect interpretation, so no single protocol suits every task. OSEM required early stopping for detection and feature stability but more iterations for mean-SUV quantification, whereas weakly regularized BSREM matched both at one setting. LesionStudio (https://github.com/qurit/LesionStudio), with our public library of realistic lesions, provides a reproducible framework for task-specific optimization across radiopharmaceuticals and scanners.

physics.med-ph↗