arXiv · 1803.06821
Development of co-located ${}^{129}$Xe and ${}^{131}$Xe nuclear spin masers with external feedback scheme
Abstract
We report on the operation of co-located ${}^{129}$Xe and ${}^{131}$Xe nuclear spin masers with an external feedback scheme, and discuss the use of ${}^{131}$Xe as a comagnetometer in measurements of the ${}^{129}$Xe spin precession frequency. By applying a correction based on the observed change in the ${}^{131}$Xe frequency, the frequency instability due to magnetic field and cell temperature drifts are eliminated by two orders of magnitude. The frequency precision of 6.2 $μ$Hz is obtained for a 10$^4$ s averaging time, suggesting the possibility of future improvement to $\approx$ 1 nHz by improving the signal-to-noise ratio of the observation.
Explore related subjects
Keep this discovery
T. Sato, Y. Ichikawa, S. Kojima, C. Funayama, S. Tanaka, T. Inoue, A. Uchiyama, A. Gladkov, A. Takamine, Y. Sakamoto, Y. Ohtomo, C. Hirao, M. Chikamori, E. Hikota, T. Suzuki, M. Tsuchiya, T. Furukawa, A. Yoshimi, C. P. Bidinosti, T. Ino, H. Ueno, Y. Matsuo, T. Fukuyama, N. Yoshinaga, Y. Sakemi, K. Asahi. 2018-03-19. Development of co-located ${}^{129}$Xe and ${}^{131}$Xe nuclear spin masers with external feedback scheme. https://doi.org/10.1016/j.physleta.2017.12.003