arXiv · 2211.05860
An RPC-PET brain scanner demonstrator: first results
Abstract
We present first results from a Positron Emission Tomography (PET) scanner demonstrator based on Resistive Plate Chambers and specialized for brain imaging. The device features a 30 cm wide cubic field-of-view and each detector comprises 40 gas gaps with 3D location of the interaction point of the photon. Besides other imagery, we show that the reconstructed image resolution, as evaluated by a hot-rod phantom, is sub-millimetric, which is beyond the state-of-the-art of the standard PET technology for this application.
Explore related subjects
Keep this discovery
Paulo Fonte, Luís Lopes, Rui Alves, Nuno Carolino, Paulo Crespo, Miguel Couceiro, Orlando Cunha, Nuno Dias, Nuno C. Ferreira, Susete Fetal, Ana L. Lopes, Jan Michel, Jorge Moreira, Américo Pereira, João Saraiva, Carlos Silva, Magda Silva, Michael Traxler, Antero Abrunhosa, Alberto Blanco, Miguel Castelo-Branco, Mário Pimenta. 2022-11-10. An RPC-PET brain scanner demonstrator: first results. https://doi.org/10.1016/j.nima.2023.168236
Cite the original work for its findings. Save a collection to share your selection of sources.