arXiv · 2308.09822
Fast Single-Quantum Measurement with a Multi-Amplifier Sensing Charge-Coupled Device
Abstract
A novel readout architecture that uses multiple non-destructive floating-gate amplifiers to achieve sub-electron readout noise in a thick, fully-depleted silicon detector is presented. This Multi-Amplifier Sensing Charge-Coupled Device (MAS-CCD) can perform multiple independent charge measurements with each amplifier; measurements with multiple amplifiers can then be combined to further reduce the readout noise. The readout speed of this detector scales roughly linearly with the number of amplifiers without requiring segmentation of the active area. The performance of this detector is demonstrated, emphasizing the ability to resolve individual quanta and the ability to combine measurements across amplifiers to reduce readout noise. The unprecedented low noise and fast readout of the MAS-CCD make it a unique technology for astronomical observations, quantum imaging, and low-energy interacting particles.
Explore related subjects
Keep this discovery
Ana M. Botti, Brenda A. Cervantes-Vergara, Claudio R. Chavez, Fernando Chierchie, Alex Drlica-Wagner, Juan Estrada, Guillermo Fernandez Moroni, Stephen E. Holland, Blas J. Irigoyen Gimenez, Agustin J. Lapi, Edgar Marrufo Villalpando, Miguel Sofo Haro, Javier Tiffenberg, Sho Uemura. 2023-08-18. Fast Single-Quantum Measurement with a Multi-Amplifier Sensing Charge-Coupled Device. https://doi.org/10.1109/ted.2024.3392711
Cite the original work for its findings. Save a collection to share your selection of sources.