The DESY Digital Silicon Photomultiplier: Device Characteristics and First Test-Beam Results

Abstract

Silicon Photomultipliers (SiPMs) are state-of-the-art photon detectors used in particle physics, medical imaging, and beyond. They are sensitive to individual photons in the optical wavelength regime and achieve time resolutions of a few tens of picoseconds, which makes them interesting candidates for timing detectors in tracking systems for particle physics experiments. The Geiger discharges triggered in the sensitive elements of a SiPM, Single-Photon Avalanche Diodes (SPADs), yield signal amplitudes independent of the energy deposited by a photon or ionizing particle. This intrinsically digital nature of the signal motivates its digitization already on SPAD level. A digital SiPM (dSiPM) was designed at Deutsches Elektronen Synchrotron (DESY), combining a SPAD array with embedded CMOS circuitry for on-chip signal processing. A key feature of the DESY dSiPM is its capability to provide hit-position information on pixel level, and one hit time stamp per quadrant at a 3 MHz readout-frame rate. The pixels comprise four SPADs and have a pitch of about 70 um. The four time stamps are provided by 12 bit Time-to-Digital Converters (TDCs) with a resolution better than 100 ps. The chip was characterized in the laboratory to determine dark count rate, breakdown voltage, and TDC characteristics. Test-beam measurements are analyzed to assess the DESY dSiPMs performance in the context of a 4D-tracking applications. The results demonstrate a spatial hit resolution on a pixel level, a minimum-ionizing particle detection efficiency of about 30 % and a time resolution in the order of 50 ps.

Explore related subjects

Keep this discovery

BibTeXRIS

Finn Feindt, Inge Diehl, Karsten Hansen, Stephan Lachnit, Frauke Poblotzki, Daniil Rastorguev, Simon Spannagel, Tomas Vanat, Gianpiero Vignola. 2024-02-19. The DESY Digital Silicon Photomultiplier: Device Characteristics and First Test-Beam Results. https://doi.org/10.1016/j.nima.2024.169321

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det