Scatter correction for photon-counting detector based CBCT imaging
Abstract
Objective: The aim of this study is to validate the effectiveness of an energy-modulated scatter correction method in suppressing scatter in photon-counting detector (PCD)-based cone beam CT (CBCT) imaging. Approach: The scatter correction method, named e-Grid, which was initially applied to dual-layer flat-panel detector (DLFPD)-based CBCT imaging, was tested for its performance in PCD-CBCT imaging. Benchtop PCD-CBCT imaging experiments were conducted to verify the effectiveness of the e-Grid method. Additionally, quantitative metrics were measured from these experimental results. Main results: It was found that the use of the e-Grid method could significantly eliminate cupping artifacts caused by Compton scatter in PCD-CBCT imaging. Meanwhile, its effectiveness was observed in both low- and high-energy images, as well as for objects of varying sizes. Quantitative results showed that the e-Grid method could reduce scatter artifacts by at least 71% in low-energy images and 75% in high-energy images. Significance: It was demonstrated that the scatter correction method originally applied to DLFPD-based CBCT could also perform well in PCD-CBCT, showing that the e-Grid method has great potential for application in other spectral CBCT imaging systems.
Explore related subjects
Keep this discovery
Xin Zhang, Ting Su, Jiongtao Zhu, Hairong Zheng, Dong Liang, Yongshuai Ge. 2024-12-17. Scatter correction for photon-counting detector based CBCT imaging. https://arxiv.org/abs/2412.12862