Performance evaluation of compact plastic scintillating fiber modules for muon tomography applications
Abstract
Muon tomography is a non-destructive imaging technique that exploits cosmic-ray muons from multiple directions. Its performance critically depends on the stability, active-area coverage, and spatial resolution of position-sensitive detectors. In this work, we report on the development of four compact scintillating fiber modules, each 100 cm long and composed of two staggered layers of 1 mm diameter fibers. The fibers are read out at one end by one-dimensional silicon photomultiplier arrays with a 2 mm pitch, coupled to Citiroc 1A-based front-end electronics. The modules were characterized with cosmic-ray muons, yielding a detection efficiency above 97% and a mean spatial resolution of about 0.56 mm, with uniform response over different distances from the readout end. An imaging test of a lead block was also performed, and the reconstructed results are consistent with the blocks profile. These results demonstrate the suitability of this detector design for compact and large-area systems in muon tomography applications.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Yiyue Li, Huiling Li, Hui Liang, Cong Liu, Chenghan Lv, Hongbo Wang, Weiwei Xu. 2025-09-18. Performance evaluation of compact plastic scintillating fiber modules for muon tomography applications. https://doi.org/10.1016/j.nima.2026.171595