arXiv · 2305.09297
Calibration of a Micromegas-based Gaseous Time Projection Chamber Using Cosmic Ray Muons
Abstract
We report the calibration of a gaseous Time Projection Chamber based on Micromegas charge readout modules with cosmic ray muons, utilizing their penetrating power and relatively uniform energy deposition per unit length. Muon events were selected through track reconstruction to characterize detector performances, such as the drift velocity, electron lifetime, detector gain, and electric field distortion. The evolution of detector performances with gas purities and electric drift fields over a 50-day data-taking cycle was measured by the muon calibration method. For instance, the drift velocity degraded with gas impurities from ${3.40\pm 0.07 ~ cm/\mu s}$ to ${3.06\pm 0.06 ~ cm/\mu s}$, and then recovered with gas purification. A ${^{137}Cs}$ calibration source was also placed inside the detector as a reference for muon calibrations.
Explore related subjects
Keep this discovery
Wenming Zhang, Yuanchun Liu, Haiyan Du, Ke Han, Heng Lin, Tao Li, Lingyin Luo, Kaixiang Ni, Yunzhi Peng, Shaobo Wang, Sicheng Wen, Xiyu Yan, Zhiyong Zhang, Wenchang Zhai. 2023-05-16. Calibration of a Micromegas-based Gaseous Time Projection Chamber Using Cosmic Ray Muons. https://doi.org/10.1088/1748-0221%2F18%2F07%2Fp07038
Cite the original work for its findings. Save a collection to share your selection of sources.