Enhanced sensitivity to trace $^{238}$U impurity of sapphire via coincidence neutron activation analysis
Abstract
Sapphire has mechanical and electrical properties that are advantageous for the construction of internal components of radiation detectors such as time projection chambers and bolometers. However, it has proved difficult to assess its $\rm ^{232}Th$ and $\rm ^{238}U$ content down to the picogram per gram level. This work reports an experimental verification of a computational study that demonstrates $\gamma\gamma$ coincidence counting, coupled with neutron activation analysis (NAA), can reach ppt sensitivities. Combining results from $\gamma\gamma$ coincidence counting with those of earlier single-$\gamma$ counting based NAA shows that a sample of Saint Gobain sapphire has $\rm ^{232}Th$ and $\rm ^{238}U$ concentrations of $<0.26$ ppt and $<2.3$ ppt, respectively; the best constraints on the radiopurity of sapphire.
Explore related subjects
Keep this discovery
D. Chernyak, I. J. Arnquist, T. Daniels, S. W. Finch, L. Hissong, M. Hughes, R. MacLellan, A. Piepke, A. Pocar, R. Roshong, R. Saldanha, R. H. M. Tsang. 2025-08-06. Enhanced sensitivity to trace $^{238}$U impurity of sapphire via coincidence neutron activation analysis. https://doi.org/10.1103/zgzm-w5f4
Cite the original work for its findings. Save a collection to share your selection of sources.